| Item | Specifications |
| Furnace Chamber Size | ≥200200250mm (LengthWidthHeight)OR≥200250mm (DiameterHeight) |
| Furnace Body Opening Mode | Lifting Type |
| Temperature Controller | Standard Digital Controller |
| - Programmable based on heating/cooling rate or time |
| - Equipped with automatic constant temperature and time control functions |
| - Digital control system enables precise temperature setting and display |
| Flow Meter | Equipped with Gas Flow Meter |
| Sealing Method | Soft Material Sealing (Equipped with Water Cooling for Temperature Reduction) |
| Operating Temperature | Long-term Operating Temperature ≥1700℃ (2-hour continuous operation at 1700℃ without abnormality) |
| Temperature Control Precision | ≤±1℃ |
| Heating Rate | Adjustable from 1-10℃/min |
| Temperature Field Uniformity | ≤±5℃ (1300-1700℃, Test Report Provided) |
| Special Gas Type | Nitrogen Supported |
| Frnace Chamber Tightnessu | Vacuum Degree ≤-0.1MPa (Measured at Room Temperature) |
| High-temperature Protection | External Surface Temperature of the Machine <50℃ at Maximum Operating Temperature |
| Wire Requirement | Power Cable is Used as the Power Supply Line |
| Protection Functions | - Automatic Heating Stop Protection when Door is Opened |
| - Over-temperature, Over-pressure, Short Circuit, Leakage Protection, etc. |
| Process Confidentiality System | Password Setting Function to Protect Process Parameters |