Technical Parameters:
| Item | Specification |
|---|---|
| Rated Temperature | 1600℃ |
| Max Operating Temperature | 1600℃ |
| Long-term Working Temperature | 500–1600℃ |
| Chamber Size | L650 × W650 × H650 mm |
| Control Range | 80~1600℃ |
| Temperature Sensor | B-type thermocouple |
| Heating Element Installation | Around the chamber (front, rear, left, right) |
| Temperature Control Accuracy | ±1℃ (Integrated circuit control, no overshoot) |
| Temperature Uniformity | ≤±5℃ within effective heating zone; measured by 5 thermocouples after holding at 1200℃ for 1 hour |
| Heating Rate | Adjustable, maximum 20℃/min |
| Cooling Rate | 1000–500℃: 5–10℃/min; 500–300℃: 2–5℃/min; Eurotherm controller, min. 0.5℃/h |
| Heating Element | MoSi₂ heating rod |
| Furnace Body | CNC machined, polished, pickled, phosphated, powder coated and high-temperature baked; two-tone design, anti-oxidation, acid & alkali resistant, high temperature resistant, easy to clean |
| Body Structure | Air-cooled double-layer furnace shell with air guide baffles, circulating cooling for conductive strips to prevent high-temperature oxidation |
| Door Opening Type | Bottom hinged door, hydraulic lifting |
| Sealing between Loading Platform & Furnace Mouth | Step soft sealing with elastic high-temperature sealing strip to absorb thermal expansion of refractory materials |
| Refractory Material | Vacuum formed high-purity lightweight alumina material; lightweight alumina bubble board for furnace mouth and bottom. Low heat storage, thermal shock resistant, crack & slag free |
| Insulation Material | 3-layer insulation: aluminosilicate fiber board, alumina fiber board, alumina polycrystalline fiber board, energy saving ≥80% vs traditional furnaces |
| Shell Surface Temp | Shell temperature <45℃ during long-time non-stop operation |
| Protection Functions | Integrated control module, dual-loop control & protection; protection against overshoot, undershoot, thermocouple break, phase loss, overvoltage, overcurrent, overtemperature, soft start; pressure interlock protection for overpressure risk |
| Control System | 10-inch true-color touch screen + PLC centralized control, data log & report export to PC; closed-loop SCR control, phase-shift / zero-cross trigger, constant voltage / constant current / constant power mode, current limiting protection |
| Water Cooling Zones | Furnace top, furnace opening, furnace door, outer shell |
| Permissible Atmosphere | Nitrogen, Argon, Oxygen, etc. |
| Vacuum Degree | -0.1Mpa (gauge), equipped with two-stage rotary vane mechanical pump |
| Gas Inlet System | 3 sets of float flowmeter + manual valve (customizable range / quantity); max slight positive pressure 0.05Mpa (gauge) |
| Exhaust System | Exhaust pipe on furnace top, inner diameter 80mm |
| Pressure Relief System | Relief port on top between liner and heating elements for discharging redundant gas |
| Display Contents | Real-time temperature, segment No., segment time, remaining time, power percentage, voltage, current, etc. |
| Push Buttons | Imported push buttons, service life>100,000 times with LED indicator |
| Temperature Curve Setting | Intelligent controller with PID / APID / MPT AI regulation, auto-tuning & self-learning, no overshoot or undershoot. 30-segment programmable control, arbitrary ramp slope, cycle jump, run, pause, stop, editable during running |
| Program Segments | 30 segments (50 segments optional); flexible allocation for multiple curves, storable and recallable |
| Communication Port | RS485, max 1200m; PC remote operation, curve editing, data logging; history curve & report storage, logging interval 1s~1h |